Series 9610 Carbon Dioxide Analyzer

The Series 9610 Carbon Dioxide Analyzer is designed to provide continuous, unattended measurement of carbon dioxide in other gases. Measuring ranges include either percentage or trace (PPM). The 9610 may be used for applications where spot monitoring of various points is of interest.
The Series 9610 Carbon Dioxide Analyzer provides continuous, unattended measurement of either percent or trace levels of carbon dioxide in other gases. The product is ideally suited for both continuous or spot monitoring requirements.
The focal point of the Series 9610 is its state-of-the-art sensor technology that features a next generation NDIR (non- dispersive infrared) measuring system. This carbon dioxide sensor incorporates closed-loop control for enhanced long-term calibration stability. The control signal is generated by means of an optical referencing element that monitors source intensity.
Read more about sample system handling for carbon dioxide analyzers
This methodology, coupled with a novel source homogenizing gas sampling chamber, yields a system with greatly improved tolerance to changes in the infrared source outputs.
Back To Top
Performance
Percent Ranges: 0-20%
Trace Ranges: 0-5,000 Parts Per Million (PPM)
Error Band:
- Percent Range: ± 0.1% or +5.0% of reading, whichever is greater
- Trace Range: ± 30 ppm or + 2% of reading,whichever is greater (error stated at 77°F and 14.7 psig)
Repeatability:
- Percent Range: ± 0.1% CO2
- Trace Range: ± 20 ppm CO2
Response Time: <35 seconds to a 63% step change @ a sample flow rate of 200 ml/minute (recommended sample flow rate)
Sensor Type: Non-Dispersive Infrared (NDIR)
Operating Humidity
Range: 0-99% RH non-condensing
Warm-up Time: Less than one minute
Sample Connections: 1/4" SS compression fitting
Sample Delivery: Pressurized sample or optional pump
Alarm Relays: Two (2) SPDT Form C contacts rated 10A (250 VAC) / 5A (100 VDC)
Display: 4 line x 20 character LCD for both O2 and CO2
Analog Outputs: Two 4-20 mADC outputs that are range configurable. Either or both outputs can be set to provide 0-20 mADC
Calibration: Single gas calibration
General Specifications
Power Requirements: 90-264 VAC, 50-60 Hz
Ambient Temperature Range: 32-122°F (0-50°C)
Sample Pressure Limits: <2.0 psig
Enclosure: NEMA 1, Powder coated, painted aluminum enclosure suitable for bench-top use.
Enclosure Dimensions: 10.75” W x 6.30” H x 13.10” D
Warranty: Two years electronics and sensor
Weight: <10 pounds (uncrated)
- RS-232 or RS-485 Serial Communications
- Sample Pumps
- Sample Filters
- Pressure Regulators
- Flow meters
- Built-in Data Logger
- Remote Sensor
- Food Processing
- Pharmaceutical Manufacturing
- Environmental Monitoring
- R&D
- Laboratory Growth Chambers
- Fermentors
- Glove Boxes and Isolators
- Incubators