Products

Trace Oxygen Analyzers
Percent Oxygen Analyzers
Safety Monitors
Carbon Dioxide Analyzers
Custom Systems

Product Selection Guide
Product Application Guide

Industries Served

Our Products are used in a wide variety of industries and applications.
Learn more...

Applications

Find out how our Products can help you. Fill out our Applications Questionnaire.

Technical Support

Please call us with your questions: 800-262-5977 or 401-333-8580

Portable or Fixed

Whether you need a portable battery operated analyzer or temperature controlled explosion proof system for a fixed installation, Alpha Omega Instruments has the solution.
Learn more...

Contact Information

Alpha Omega
Instrument Corp.
40 Albion Road, Suite 100
Lincoln, RI 02865
Toll-Free: 800.262.5977
Phone: 401.333.8580
Fax: 401.333.5550

Contact Us

Series 9600 Oxygen and Carbon Dioxide Monitor

The Series 9600 Oxygen and Carbon Dioxide Monitor is designed to provide continuous, unattended monitoring of both oxygen and carbon dioxide.

 

Features Benefits
State-of-the-Art Sensor Technology High Precision Measurement
Wide Ranges Provides Greater Instrument Versatility
User Scalable Analog Outputs Better Defines the Concentrations of Interest
Rapid Speed of Response Quickly Senses Changes in Gas Composition
Dual Adjustable Alarm Setpoints Set Critical Process Limits
Advanced Digital Electronics Helps to Ensure Accurate Measuring Values
Minimum Maintenance Low Cost of Ownership
4 Line Back-lit LCD Easy to Read in a Variety of Ambient Conditions

System Description

The Series 9600 Oxygen and Carbon Dioxide Analyzer is designed to provide continuous, unattended monitoring of both oxygen and carbon dioxide. The focal point of the Series 9600 is its state-of-the-art sensor technologies. The oxygen sensor is an extended life electrochemical sensor with EES (enhanced electrolyte system). This sensor provides exceptional accuracy and stability. EES retards passivation of the sensor anode by allowing the products of oxidation to dissolve in the electrolyte. In effect, the sensor is renewed continuously resulting in a significant increase in sensor life. In addition, the enhanced mechanical design of the sensor helps to ensure long life and virtually eliminates leakage of electrolyte, a nagging (and expensive) problem associated with other types of sensors.

Learn about about sample system handling for oxygen and carbon dioxide analyzers

Alpha Omega Instruments has complemented its oxygen measuring capabilities with a next generation NDIR (non-dispersive infrared) sensor. The carbon dioxide sensor incorporates closed-loop control for enhanced long-term calibration stability. The control signal is generated by means of an optical referencing element that monitors source intensity. This methodology, coupled with a novel source homogenizing gas sampling chamber, yields a system with greatly improved tolerance to changes in the infrared source outputs.

Specifications

OXYGEN
Ranges:

  • Percent: 0-100

Error Band: ±1% of full scale
Response Time: 90% of full scale in < 20 seconds
Sensor Type: Long-life Electrochemical Sensor with EES
Operating Humidity Range: 0-99% RH non-condensning
Gas Sample Connections: 1/4" SS compression fitting
Sample Delivery: Pressurized sample or optional pump
Alarm Relays: Two (2) SPDT Form C contacts rated 10 A (250 VAC) / 5A (100 VDC) User configurable to alarm on either O2 or CO2 levels
Display: 4 line x 20 character LCD for both O2 and CO2
Analog Outputs: Two 4-20 mADC outputs that are range configurable. Either or both outputs can be set to provide 0-20 mADC

CARBON DIOXIDE
Ranges:

  • Percent: 0-20%
  • Trace: 0-5,000 Parts Per Million (PPM)

Error Band:

  • Percent Range: + 0.1% or + 5.0% of reading, whichever is greater
  • Trace Range: ± 30 ppm or + 2% of reading, whichever is greater (error stated at 77°F and 14.7 psig)

Response Time: <35 seconds to 63% of step change @ a sample flow rate of 200ml/minute (recommended sample flow rate)
Sensor Type: Non-dispersive infrared (NDIR)
Operating Humidity Range: 0-99% RH non-condensing
Gas Sample Connections: 1/4" SS compression fitting
Sample Delivery: Pressurized sample or optional pump
Alarm Relays: Two (2) SPDT Form C contacts rated 10 A (250 VAC) / 5A (100 VDC) User configurable to alarm on either O2 or CO2 levels
Display: 4 line x 20 character LCD for both O2 and CO2
Analog Outputs: Two 4-20 mADC outputs that are range configurable. Either or both outputs can be set to provide 0-20 mADC

GENERAL SPECIFICATIONS
Power Requirements: 90-264 VAC, 50-60 Hz
Ambient Temperature Range: 40-100°F (5-38°C)
Sample Pressure Limits: <2.0 psig.
Enclosure: NEMA 1, Power coated, painted aluminum enclosure suitable for bench-top use
Enclosure Dimension: 10.76" W x 6.3" H x 13.10" D
Warranty: Two years electronics and sensors
Weight: <10 pounds (uncrated)
Back To Top

Options/Accessories

  • RS232 & RS-485 Serial Communications
  • Sample Pumps
  • Sample Filters
  • Pressure Regulators
  • Flow Meters
  • Build-in Data Logger
  • Remote Sensors

Back To Top

Applications

  • Food Processing
  • Pharmaceutical Manufacturing
  • Environmental Monitoring
  • Chemical/Petrochemical
  • R&D
  • Laboratory Growth Chambers
  • Glove Box/Dry Box
  • Incubators
  • Fermentors
  • Isolators